LS-1866
Differential pressure | Minimum display | 0.1Pa(Standard) |
---|---|---|
Display range | ±999Pa | |
Excess pressure on sensors | 5Mpa | |
Leak volume display range | 0.00~±999mL/min(Auto range) | |
Alarm setting range | 0~±999 | |
Number of Channels | 16 | |
Timer setting range | 0~999.8s (Setting is infinite) | |
Power Source | AC100~240V±10%, 50/60Hz, 70VAmax | |
Test pressure source | Uses clean air. The original pressure must be sufficiently higher than the test pressure. | |
Pilot pressure source | Applies 400 to 700kPa of clean air. | |
Pipe connection diameter | Rc1/4 (air pressure source, workpiece piping, master piping), 6mm one-touch joint (pilot air source) (the pilot air source of pressure range M, L are connected to the air pressure source with a 6mm tube.) | |
Transmission function | Serial port RS-232C (printer mode setting possible) | |
Ambient temperature/humidity | 5 to 40°C/80%RH or less, however no dew condensation | |
Weight | Approx. 6.3kg (H type: approx. 8kg) | |
Standard fittings and accessories | (Fitting to main body) regulator, oil mist separator, filter regulator for pilot pressure (Accessories) Quick mounting bracket, interface connector, power cord, inspection report, user manual, traceability documents |
A, B and D in the model notation are mandatory.
A | Pneumatic | Intelligent 1 Pneumatic Circuit | A1 | Large flow circuit with great sensor protection features | |||
---|---|---|---|---|---|---|---|
Micro Volume Circuit | AS01 | For the parts whose volume is approx. 10 mL or smaller and the leak specifications are quite small. |
|||||
B | Test Pressure Range |
Medium pressure | M | Test pressure range 50 to 800 kPa (PS 1 MPa, Regulator 0.8 MPa) |
|||
Low pressure | L | Test pressure range 10 to 95 kPa (PS 100 kPa, Regulator 200 kPa) |
|||||
Vacuum pressure | V | Test pressure range -5 to -100 kPa (PS -100 kPa, Regulator -100 kPa) |
|||||
High pressure | H10 | Test pressure range 100 to 990 kPa (PS 1 MPa, Regulator 1 MPa) |
|||||
High pressure | H15 | Test pressure range 0.5 to 1.5 MPa (PS 2 MPa, Regulator 2 MPa) |
|||||
C | Option | External Exhaust Valve Ready *1 | G | External Exhaust Valve is sold separately. | |||
Blow Orifice Ø 1.2 *2 | F1 | Standard orifice changed from Ø 3.2 to Ø 1.2 | |||||
DPS 10 kPa | D4 | Sensor range: ±10 kPa Resolution: 1 Pa | |||||
D | Power Cable | VA | 125 V (3 m) | ||||
VE | 250 V (2 m) | ||||||
VK | 250 V (2 m) (Only for Chinese customers) |
*1 Standard specification. Not selectable for Vacuum (V).
*2 Required for AS01. Also, select when the test pressure is -10 to -4 kPa.