AF-2400

  • AF-2400-120911

Air flow tester has a compact body with full of useful functions.

Equips the mass flow sensor and all necessary measurement circuit is built inside of its compact body. You just need to press the start button and it will automatically carry out the test and make judgment.

Edge & Function



Features

*Full-dot LCD with back light

*Easy-to-operate "Basic" setup mode is available

*Maximum of 1000 data can be stored as a bar graph. Also, the compensation sampling data available in order to optimize the quality control.

*Wide range of flow is available from 500mL/min to 100L/min. Traceability available.

Option

*Bypass circuit ready
Effective for large volume workpiece under the test. Extra pressurization passage at the time of CHG stage by built-in Bypass valve.

*External exhaust valve ready (Exhaust valve is sold separately)
Exhaust by the external unit. Exhaust valve would prevent oil/contaminants from the workpiece under the test.

Functions
Display Full-dot LCD with back light
Flow sensor Mass flow sensor
Charge Hold Keep pressurization infinitely
Limit setting Two high level (HH/Hi) NG and one low level (Lo) NG setting
Noise reduction Repeating the test if the result falls in Hi-NG to HH-NG range
Set mode Basic: Quick setup of basic parameter
Advanced: Detailed and customized settings
Compensation Automatic compensation after figuring out the correct drift amount
Exhaust Interference Prevention Exhaust timing can be externally controlled
I/O NPN-PNP dual type
Sensor protection Air blow at every exhaust stage. In every test, the cleaning of the pneumatic circuit is automatically carried out.
Bar graph display Maximum of 1000 data can be stored as a bar graph. Also, the compensation sampling data available in order to optimize the quality control.
Detection judgment mode Judgment either at the end of DET stage or when the measurement value exceeds the limit
Conversion temperature Selection either 20℃ or 0℃
Digital filter Filter constant (0 to 99) to average the readings and produce highly stable measurement
Counter The number of Total, Pass, and Fail
Password change Any 4 digit numbers can be chosen as own password

Spec



Flow range
(Display value)
500mL/min (000~500)

2L/min (0.000~2.000)

5L/min (0.00~5.00)

20L/min (00.00~20.00)

50L/min (00.0~50.0)

100L/min (000.0~100.0)

Flow sensor Mass flow sensor
Pressure media Air
Accuracy ±1.5% of F.S. ±1 digit (At specified pressure)

±3% of F.S. ±1 digit (At non-specified pressure)

Specified test pressure range Micro (L03) : 10~30kPa

Low (L1) : 30~80kPa

Medium (M) : 30~700kPa

Vacuum (V) : -10~-70kPa

Pressure proof 1.5MPa
Channels 32ch (0~31)
Timer setting 999.9 sec (min 0.1sec)
Power supply AC100~240V±10%, 50/60Hz,100VA max
(Power cable included is for AC125V only)
Pressure source Clean and dry air. Pressure source should be higher than the test pressure
Connecting port Rc 1/4 (Pressure source, Pilot pressure)

Rc 1/2 (WORK port)

CPU 16 BIT CPU
Ambient temperature 5~45 ℃
Humidity Max. 80%RH and no condensation
Weight Approx. 12kg
RS-232C Format T Format  Fixed length output Output for judgment signal and flow rate data
Serial printer output Printer format (Data output) Judgment result, Leak rate, Compensation value, Test pressure, WK#, CH#, Time etc.

(Setting output) Setting parameter of all screen

Flow rate display L/min、mL/min、L/s、mL/s、L/h、m3/h、

USP (User Span)

Pressure Unit kPa、MPa
Standard and accessories Quick connection fittings, I/O connector, Power cable (3m), Inspection record, and Operation manual

Classification



AF-2400_spec

Pressure range

CodePressure range CodePressure range

Micro L03 10~30kPa Medium M 30~700kPa
Low L1 30~80kPa Vacuum V -10~-70kPa

 

Option code

Code Option Description
B Bypass circuit ready Effective for large volume workpiece under the test. Extra pressurization passage at the time of CHG stage by built-in Bypass valve.
G External exhaust valve ready (Exhaust valve is sold separately) Exhaust by the external unit. Exhaust valve would prevent oil/contaminants from the workpiece under the test.
N6 Cylinder control signal Output available from the start till just before the end signal. Can be used for clamping signal
P Speed controller for regulator relief Speed controller used for stabilizing the pressure adjustment

Flow range

Pressure range
Flow range Positive pressure Negative pressure
500mL/min Micro (L03) : 10~30kPa
Low (L1) : 30~80kPa
Medium (M) : 30~700kPa
Vacuum (V) : -10~-70kPa

Up to 20L/min range

2L/min
5L/min
20L/min
50L/min
100L/min

Specified test pressure range

Specified test pressure range Micro (L03) : 10~30kPa
Low (L1) : 30~80kPa
Medium (M) : 30~700kPa
Vacuum (V) : -10~-70kPa

 